Expandable polyurethane stent valve for transcatheter implantation in children suffering from heart valve disease; results of physical, hydrodynamic and experimental tests

Authors: Miguel Angel Maluf, Mariana G. Massei, Rogerio Orbetelli, Miko Obradovic, Christian Bader, Christoph Klaus, Evandro Drigo, Rosa Sa, Andre

Gimenes. Fernando Zanoni, Caio Sabino, Ovandir Bazan, Jayme Ortiz Sao Paulo Federal University, Brazil

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

Introduction: Transcatheter valves manufactured using biological tissue, as the essential structural component, can be induced to: mechanical degradation after crimping and early calcification in pediatric patients. Objectives: Manufacture and successful tests of one expandable polyurethane stent valve, may reduce the repeated operations of valve replacement, during the growing children. Material / Methods: I)Physical testing: Prostheses were submitted to universal testing in machine EMIC and a computer with Tesc software, able to generate graphs of force versus deformation (stretching). II)Hydrodynamic testing: Prostheses with diameters from 12 to 22 mm, were submitted to pulsatile physiological flow and stress conditions. III)Experimental implants in sheep: Ten sheep was submitted to prosthesis implant, by trans catheter technique in pulmonary position. In Group A: Four sheep w/ 20 kg, expanded up to 22mm. Results: Physical and Hydrodynamic testing of Polyurethane strip removal of stent valve, before and after undergoing to 30 minutes crimping, showed preservation of properties of resistance and elasticity elongation. In vitro durability was proven for >15 years. Eight sheep, were submited to 3D echo study, performed in the 6th month of f.-up, showed: there was no significant transvalvular gradients and trivial regurgitation in 3 cases. Histologic, radiologic and electron microscopic study of the first prosthesis shows: integrity of structure and free of calcification. Seven survival sheep are well, after 24th months of follow-up. Conclusion: Expandable polyurethane stent valve, with special design for implant and expansion in growing children, has experimental satisfactory hemodynamic performance and durability in vitro and in vivo tests. Calcification and structural changes were not observed. In the next step, clinical studies are be planned.

Subjects

Cardiology

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.

Try:

Similar by topic

More from this journal

The newest in International Journal of Cardiovascular Research.

All its articles

Conferences in this field

Upcoming conferences in this field.

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalInternational Journal of Cardiovascular Research
  • SubjectCardiology
  • BylineArticles whose byline prints “Miguel Angel Maluf”
  • BylineArticles whose byline prints “Mariana G. Massei”
  • BylineArticles whose byline prints “Rogerio Orbetelli”
  • BylineArticles whose byline prints “Miko Obradovic”
  • BylineArticles whose byline prints “Christian Bader”
  • BylineArticles whose byline prints “Christoph Klaus”
  • BylineArticles whose byline prints “Evandro Drigo”
  • BylineArticles whose byline prints “Rosa Sa”

For authors and organisers

Publish, present and keep up with your field

Find the right journal for your paper and submit your manuscript.

Photo: Pexels