Article · 29 Dec 2025
Conference abstractPublished 7 Nov 2016
Comparison of conventional vs. software-based methods for rapid arc SRS QA
Vantage Oncology Center for Radiation Therapy of Beverly Hills, USA
Abstract
Select any words in the title or abstract, or press this and choose a sentence.
Aim of this study is to evaluate a SRS rapid arc QA method using the portal dosemeter as a viable QA tool and comparison of its results to the conventional physicâÂ?Â?s QA methods, the film and ion chamber measurements. Both absolute and relative QAsâÂ?Â? results have been studied in this study. Ion chambers have a high dependancy on their setup and more specifically on their placement within the radiation field. Large variations in their readings can stem from a minute setup deviation. This is especially accentuated in the case of a the high dose gradient areas of a small field SRS field. The results are very dependant on this placment accuracy. On the other hand, the outcome is limited to one absolute measurement data per placement. The same can be said about film dosimetry and its own uncertainties such as developer conditions, film batch and the film development procedure as a whole. Similar to the ion chamber setup, one can expect one relative measurement data out of each film setup. We are presenting a portal dosimetry based QA tool that will not have similar uncertainties but on the other hand, produces range of measurment data per exposue, based on acctual delivery plan not hypothetical results. We are comparing its performance to the conventional methods of the ion chamber and film based measurements.
Subjects
Biochemistry & BiotechnologyAsk about this article
Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.
More from this journal
The newest in Journal of Bioengineering & Biomedical Science.
Article · 29 Dec 2025
3D Bioprinting: Personalized Medicine’s Future Frontier
Article · 29 Dec 2025
Microfluidics: Revolutionizing Point-of-Care Diagnostics Speed and Accuracy
Conferences in this field
Upcoming conferences in this field.
Follow
New articles from what you follow go into your email digest, once you agree to alert emails.
- JournalJournal of Bioengineering & Biomedical Science
- SubjectBiochemistry & Biotechnology