Scholars Central

Research articlePublished 2014

Novel Split Chest Tube Improves Post-Surgical Thoracic Drainage

Authors: Albert H Olivencia-Yurvati, Brandon H Cherry, Hunaid A Gurji, Daniel W White, J Tyler Newton, Gary F Scott, Besim Hoxha, Terence Gourlay, Robert T. Mallet

Clinical and Experimental Cardiology

Volume 5 · Issue 6 · Pages 1–4

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

Objective: Conventional, separate mediastinal and pleural tubes are often inefficient at draining thoracic effusions. Description: We developed a Y-shaped chest tube with split ends that divide within the thoracic cavity, permitting separate intrathoracic placement and requiring a single exit port. In this study, thoracic drainage by the split drain vs. that of separate drains was tested. Methods: After sternotomy, pericardiotomy, and left pleurotomy, pigs were fitted with separate chest drains (n=10) or a split tube prototype (n=9) with internal openings positioned in the mediastinum and in the costo-diaphragmatic recess. Separate series of experiments were conducted to test drainage of D5W or 0.58 M sucrose, an aqueous solution with viscosity approximating that of plasma. One litre of fluid was infused into the thorax, and suction was applied at -20 cm H2O for 30 min. Results: When D5W was infused, the split drain left a residual volume of 53±99 ml (mean value ± SD) vs. 148 ± 120 for the separate drain (P = 0.007), representing a drainage efficiency (i.e. drained vol/[drained + residual vol]) of 95 ± 10% vs. 86 ± 12% for the separate drains (P = 0.011). In the second series, the split drain evacuated more 0.58 M sucrose in the first minute (967 ± 129 ml) than the separate drains (680 ± 192 ml, P<0.001). By 30 min, the split drain evacuated a similar volume of sucrose vs. the conventional drain (1089 ± 72 vs. 1056 ± 78 ml; P = 0.5). Residual volume tended to be lower (25 ± 10 vs. 62 ± 72 ml; P = 0.128) and drainage efficiency tended to be higher (98 ± 1 vs. 95 ± 6%; P = 0.111) with the split drain vs. conventional separate drains. Conclusion: The split chest tube drained the thoracic cavity at least as effectively as conventional separate tubes. This new device could potentially alleviate postoperative complications.

Subjects

Cardiology

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.

Try:

More from this journal

The newest in Clinical and Experimental Cardiology.

All its articles

Conferences in this field

Upcoming conferences in this field.

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalClinical and Experimental Cardiology
  • SubjectCardiology
  • BylineArticles whose byline prints “Albert H Olivencia-Yurvati”
  • BylineArticles whose byline prints “Brandon H Cherry”
  • BylineArticles whose byline prints “Hunaid A Gurji”
  • BylineArticles whose byline prints “Daniel W White”
  • BylineArticles whose byline prints “J Tyler Newton”
  • BylineArticles whose byline prints “Gary F Scott”
  • BylineArticles whose byline prints “Besim Hoxha”
  • BylineArticles whose byline prints “Terence Gourlay”

For authors and organisers

Publish, present and keep up with your field

Find the right journal for your paper and submit your manuscript.

Photo: Pexels