Amount of trace moisture in process gases supplied to semi-conductor industry is considered to be more and more crucial. To satisfy the demand for measurement of trace moisture, instrument companies have developed or manufactured a variety of analyzers. One of those analyzers which is based on Cavity Ring-Down Spectroscopy (CRDS) technique has been indicated with better performance due to its superior response time and stability from 10 nmol/mol (ppb) to 10 Ã?¼mol/mol (ppm) of moisture in nitrogen gas. However, using gas cylinders as sources of trace moisture is not a proper way to evaluate the performance of analyzers, as substantial adsorption of water molecules on the inner surface of cylinder. Instead, this study established the dynamic trace-moisture generator which employs the diffusion method to produce stable concentration of water vapor once it reaches thermal equilibrium and adsorption-desorption equilibrium. With the dynamic trace-moisture generator, a commercial CRDS analyzer was under test on real-time and the results shows that the operating wavelength of laser might not be adjusted perfectly to the absorption band of water molecule. Therefore, the measurement stability of analyzer from direct reading could be improved by roughly a factor of 3, relative standard deviation from 1.96% to 0.64% at a level of 10 Ã?¼mol/mol through a tiny region spectrum scanning. This study demonstrates a feasible manner that generates trace moisture dynamically to evaluate the performance of commercial analyzer. It could be anticipated that a further improvement of measurement based on CRDS is to resolve the molecular absorption spectrum more precisely.
Subjects
ChemistryAsk about this article
Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.
Similar by topic
Conference abstract
Measuring water vapor permeation for food package materials using cavity ring-down spectroscopy
Review article · 25 Jan 2013
Cavity Ring Down Laser Absorption Spectroscopy For Evaluation Drug Molecule
Research article · 2 Jan 2023
Plasmon-Induced Flexibility and Refractive Index Measurement in a Sensor Designed by a Cavity, Two Rings, Two Teeth and Two Plasmonic Waveguides
Conference abstract · 13 May 2019
Hybrid ring /linear cavity fiber laser operating at 1064nm band for sensing
More from this journal
The newest in Chemical Sciences Journal.
Brief report · 29 Dec 2025
Green Chemistry: Sustainable Synthesis For Future Pharmaceuticals
Brief report · 29 Dec 2025
Cheminformatics: Data-Driven Chemistry's Transformative Impact
Opinion · 29 Dec 2025
Polymers: Engineering Biomedicine's Future Through Innovation
Conferences in this field
Upcoming conferences in this field.
Follow
New articles from what you follow go into your email digest, once you agree to alert emails.
- JournalChemical Sciences Journal
- SubjectChemistry
- BylineArticles whose byline prints “Chia-Wei Chang”
- BylineArticles whose byline prints “Yi-Cyun Yang”