Value of a Novel 7-Day Patch Electrocardiography on Monitoring Cardiac Arrhythmias

Xu Z, Zhao Y, Yan J, Chen X

Abstract

Objective: With the development of ECG monitoring systems, the diagnosis of cardiac arrhythmias is improvedsignificantly. Here, we introduced our experience with another new long-term ECG patch device (Smartpatch II) formonitoring the cardiac arrhythmia.Method: This device was applied to patients complained with a history of chest pain, palpitation or syncope andadmitted to the 3 medical centers between April 13, 2017 and October 18, 2017. Patients wore both 24-hour Holtermonitor and the Smartpatch II on the first day. On the second day, the 24-hour Holter monitor was removed andpatients continued to wear the Smartpatch II up to 7 days. All ECGs were analyzed by 2 experienced independentobservers.Results: A total of 101 patients (mean age: 57.5 ± 15.4 years, 52.5% male patients, body mass index: 24.0 ± 2.9kg/m2) were included. Over the first 24 hours, Both the Smartpatch II and Holter monitor documented 36 events. Thesensitivity and specificity of Smartpatch II was 100% (95% CI, 90.3%, 100.0%) and 100.0% (95% CI, 94.5%,100.0%) respectively when compared with Holter monitor on the first day. Over the total 7 days of monitoring,Smartpatch II detected significantly more arrhythmic events than one day recording (59 (58.4%) versus 36 (35.6%),p<0.001).Conclusion: Over a total wearing time of 7 days, the patch device detected more cardiac arrhythmia than 24-hour monitoring.