25 Hydroxy vitamin D quantitative analysis by LC/MS-assay standardization surprises

Authors: Eduard Rogatsky

Wadsworth Center, USA

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

Vitamin D deficiency is a widespread clinical problem and has been associated with many adverse health outcomes. Analysis of Vitamin D2 (ergocalciferol) and D3 (cholecalciferol) and their major metabolites 25(OH)D2 and 25(OH)D3 has become a high priority topic in clinical analysis. Currently a variety of LC/MS methods have been developed to support vitamin D analysis. These LC/MS methods utilize different transitions, ionization modes, sample preparation strategies, mobile phases and columns. In LC/MS analysis of 25 OH Vitamin D, dehydration (water loss) is the major side reaction. Comparing acetonitrile to methanol, which are typically used as mobile phases for LC separation, acetonitrile does not support hydrogen bond formation; therefore, proton-induced water elimination in-source becomes a major side-reaction, especially given the low pH of the mobile phase and positive mode electrospray and APCI ionization. MeOH, in contrast, supports hydrogen bond formation with the 25(OH)D2 and 25(OH)D3 hydroxyl groups. This efficiently âÂ?Â?shieldsâÂ? most of hydroxyl groups by hydrogen bonding, and protects against protonation and resultant water elimination. We found that quantitation of the 25(OH)D from its [M+H]+, âÂ?Â?intactâÂ? precursor ion, is temperature invariant. In contrast, quantitation using the in-source dehydrated precursor (parent) ion, leads to increased sensitivity with a rise in temperature, due to its better ionization efficiency at higher temperatures. Since droplets evaporation region can vary with mass spectrometer hardware design, ratio between intact [M+H]+ and dehydrated precursor can be unpredictable. We also noticed that degree of dehydration is concentrationdependent. Chromatographic separation between analyte and its deuterated internal standard might cause different levels of analyte and internal standard dehydration and resulted in quantitative error.

Subjects

Chemistry

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.

Try:

Similar by topic

More from this journal

The newest in Journal of Chromatography & Separation Techniques.

All its articles

Conferences in this field

Upcoming conferences in this field.

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalJournal of Chromatography & Separation Techniques
  • SubjectChemistry
  • BylineArticles whose byline prints “Eduard Rogatsky”

For authors and organisers

Publish, present and keep up with your field

Find the right journal for your paper and submit your manuscript.

Photo: Pexels