Case reportPublished 9 Aug 2016
Sixty inappropriate defibrillation shocks in several days due to lead displacement
Abstract
Select any words in the title or abstract, or press this and choose a sentence.
Background: Implantable cardioverter-defibrillators (ICDs) are active cardiac implants for immediate treatment of ventricular arrhythmias. They provide life-saving therapy, but may also deliver inappropriate therapy. Methods and Results : The present report describes a rare occurrence of many inappropriate ICD shocks because of lead displacement shortly after implantation. A 57-year-old overweight man received 353 antitachycardia pacings and 60 shocks of 41 J in total, all within 13 days. The device records show a decrease in the sensed R-waves and pacing impedance fluctuation. Conclusion: The ventricular electrogram signals recorded and shock channels were typical for lead rubbing.