Design of a Hydrodynamic Performance Bench for Ventricular Assist Devices

Authors: Bruno Santos, Tarcisio Leao, Eduardo Bock, Evandro Drigo, Jeison Fonseca, Aron Andrade

Journal of Bioengineering & Biomedical Science

Volume 11 · Issue 12 · Pages 1–8

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

This work presents the design of a hydrodynamic performance bench (HPB) of Ventricular Assist Devices (VADs) for evaluations of developed prototypes. VADs are used for the treatment of patients with Congestive Heart Failure (CHF), either as a bridge to recovery, for transplantation or as destination therapy. HPB is required for the performance evaluation of VADs that are under development in Brazil. The performance evaluation of a VAD takes into account the rotation speed [rpm], flow rate [l/min], pressure [mmHg] and power [W]. The project consists in choosing an actuator and transducers from HPB, and developing the mechanical components, supervisory control and modules. The HPB mechanical components were designed in SolidWorks® (Dassault Systemes, Vélizy-Villacoublay, France) following rapid prototyping and built by additive manufacturing. The HPB supervisory program was developed in graphic language through the Labview® program (2015, National Instruments, Austin, USA) and implemented in a development and application platform with data acquisition system (PXIe-8840, National Instruments, Austin, USA). The manipulated variable of the supervisory system is the motor speed [rpm] acting via an Escon 50/4 EC-S power controller (Maxon Motor, Sachseln, Switzerland) on the BLDC EC45 N339281 motor (Maxon Motor, Sachseln, Switzerland). The controlled variable is the “VAD flow” [l/min] provided by an HT-110 flow transducer. A particle image velocimetry (PIV) module was developed for flow analysis. All process data are stored in spreadsheets for later consultation. HPB was able to assist in a satisfactory way in the analysis of the developed VADs prototypes, demonstrating the technical success of the project.

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.

Try:

Similar by topic

Conferences in this field

Upcoming conferences in this field.

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalJournal of Bioengineering & Biomedical Science
  • SubjectBiochemistry & Biotechnology
  • BylineArticles whose byline prints “Bruno Santos”
  • BylineArticles whose byline prints “Tarcisio Leao”
  • BylineArticles whose byline prints “Eduardo Bock”
  • BylineArticles whose byline prints “Evandro Drigo”
  • BylineArticles whose byline prints “Jeison Fonseca”
  • BylineArticles whose byline prints “Aron Andrade”