Scholars Central

Research articlePublished 1 Dec 2020

The pace maker/reel-syndrome: rare cause of pace maker failure

Authors: Abubakar Hilal

Current Trends in Cardiology

Issue S-2020 · Pages 14

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

Pacemaker lead dislodgment is a rare and disturbing cause of pacemaker malfunctioning. The displacement results by inadvertent physical manipulation of pacemaker pulse generator by the patient. This further proceeds into lead coiling, dislocation and pacemaker failure. In this study, a case report of an 80-year ischemic male who came to the hospital with dizziness, syncope and Brady arrhythmias is described. On examination, it was found that he had been previously implanted with PPM, 12 lead EKG was now showing third degree AV block. An urgent CXR was performed which showed ventricular lead retraction. Patient had accidently manipulated pacemaker resulting into subsequent coiling of the lead. On immediate basis, a TPM wire was inserted and set at 60bpm. Ventricular lead was repositioned and pulse generator was fixed to pectoral muscle. Patient was discharged successfully and remained stable after. Pacemaker lead repositioning is a life saving treatment along with proper suture fixation of the lead and pulse generator within patient’s pocket.

Subjects

Cardiology

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.

Try:

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalCurrent Trends in Cardiology
  • SubjectCardiology
  • BylineArticles whose byline prints “Abubakar Hilal”