Editorial · 27 Apr 2026
Research articlePublished 25 Jul 2016
Achieving Consistent, Homogeneous, Dark Fat Suppression on Bilateral Breast MRI at 3.0 Tesla in the Clinical Setting
Department of Radiology and Biomedical Imaging University of California, San Francisco School of Medicine, University of California, USA; Siemens Medical Solutions USA, San Francisco, USA; Kaiser San Rafael Medical Center, San Rafael, USA
Abstract
Select any words in the title or abstract, or press this and choose a sentence.
Purpose: Effective fat suppression is a fundamental aspect of diagnostic quality breast MR for cancer screening but is challenging at 3T due to higher B0 and B1 magnetic field inhomogeneities compared with 1.5T. The purpose of this study is to describe a technique to achieve consistent homogeneous, dark fat suppression for T1 breast MR imaging at 3T for clinical breast cancer screening and evaluation. Methods: This was an IRB approved, HIPAA compliant study. Over 100 clinical breast MRI patients were scanned from May 2010 through October 2010, as part of routine clinical care, on a wide-bore 3T Magnetom Verio (Siemens Medical Solutions), One pre-contrast and two post-contrast axial VIBE scans were acquired for dynamic T1 imaging of the bilateral breasts. A longer TE of 4.9 ms was chosen empirically for optimal fat saturation. Two MQSA-certified breast radiologists, independently reviewed 20 consecutive MR studies. MR images were assessed for homogeneity of fat suppression and degree of fat suppression, both on a 5-point scale. A kappa coefficient was calculated for inter-reader agreement for both homogeneity of fat suppression and degree of fat suppression. Results: Overall, we found minimal unsuppressed fat and overall dark gray fat, indicating a high degree of fat suppression on 3T MR. Readers rated 3T fat suppression homogeneity as having minimal to no inhomogeneity in 80-90% of the examinations, with moderate agreement, (K=0.62, p<0.003), and fat suppression degree was rated having dark to very dark fat for all examinations with perfect agreement (K=1.0, p<0.001). Conclusion: Homogeneous, dark fat suppression on bilateral axial breast MRI can be obtained at 3.0T consistently in the clinical setting with minimal unsuppressed fat.
Subjects
Radiology & ImagingAsk about this article
Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.
More from this journal
The newest in Journal of Radiology.
Editorial · 30 Dec 2025
Radiology Safety: Culture, Communication, and Continuous Improvement
Editorial · 30 Dec 2025
Medical Image Quality: Objective, Subjective, and Advanced Assessment
Follow
New articles from what you follow go into your email digest, once you agree to alert emails.
- JournalJournal of Radiology
- SubjectRadiology & Imaging
- BylineArticles whose byline prints “Heather I Greenwood”
- BylineArticles whose byline prints “Vignesh A Arasu”
- BylineArticles whose byline prints “Vibhas S Deshphande”
- BylineArticles whose byline prints “Dorota J Wisner”
- BylineArticles whose byline prints “Gerhard A Laub”
- BylineArticles whose byline prints “Nola M Hylton”
- BylineArticles whose byline prints “Bonnie N Joe”