ATP Measurement in the Last Rinse Water of Automated Washer-Disinfectors: The Added Value of Every Load Monitoring

Authors: Francesco Tessarolo, Lorna Valentinotti, Giuliano Mariotti, Andrea Ziglio, Marta Rigoni, Giandomenico Nollo, Josephus PCM van Doornmalen Gomez Hoyos

1Department of Industrial Engineering, University of Trento, Mattarello Trento, Italy; 2Healthcare Research and Innovation Program (IRCS-PAT), Bruno Kessler Foundation, Trento, Italy; 3Department of Sterilization, Rovereto Hospital, Azienda Provinciale per i Servizi Sanitari, Trento, Italy; 4Direction of Rovereto Hospital, Azienda Provinciale per i Servizi Sanitari, Trento, Italy; 5Department of Applied Physics, Eindhoven University of Technology, MB, Eindhoven, The Netherlands

Journal of Clinical Infectious Diseases & Practice

Volume 1 · Issue 1 · Pages 1–6

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

Objective: To assess whether monitoring every loads of automated washer-disinfectors (AWDs) by testing the last rinse water with Adenosine Triphosphate (ATP) would have an added value to document effective and reproducible AWD processes. Methods: Every load monitoring (ELM) of the last rinse water with ATP was performed on three AWDs at the sterilization department of a 329 acute-bed hospital in Italy. No change were made in the routine procedures, except the last rinse water of a process was collected with a water sampler and tested on ATP. Data from negative controls and empty cycles were also obtained for comparison. ATP values were analysed according to AWD, load, and process characteristics. Results: A total of 435 ATP tests were performed, including 74 negative, 74 empty runs and 287 regular load tests. ATP values from regular loads ranged between 2 and 452 RLUs and 1.3%, 0.7%, 0.3% out of the 287 measurements was above 200, 300, and 400 RLUs, respectively. ATP values from negative tests and empty runs differed significantly (p<0.001) from the regular loads with a median value [I quartile; III quartile] of 4[2;5], 5[3;7], and 12[7;22] RLU, respectively. Analysis of ATP values from regular loads grouped by load pre-treatment, AWD washing program, load material or load amount showed significant differences between groups. No significant differences in ATP data were found among AWDs. Conclusion: The measurement of ATP in the last rinse water of AWDs showed that characteristics and quantities of the load relate to the ATP readings. The test is advised as an informative additional test to visual inspection for monitoring every load of an AWD.

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.

Try:

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalJournal of Clinical Infectious Diseases & Practice
  • SubjectInfectious Diseases
  • BylineArticles whose byline prints “Francesco Tessarolo”
  • BylineArticles whose byline prints “Lorna Valentinotti”
  • BylineArticles whose byline prints “Giuliano Mariotti”
  • BylineArticles whose byline prints “Andrea Ziglio”
  • BylineArticles whose byline prints “Marta Rigoni”
  • BylineArticles whose byline prints “Giandomenico Nollo”
  • BylineArticles whose byline prints “Josephus PCM van Doornmalen Gomez Hoyos”