Research article · 6 Oct 2026
Research articlePublished 2013
The Relationship between Corneal Biomechanics and Corneal Shape in Normal Myopic Eyes
Abstract
Select any words in the title or abstract, or press this and choose a sentence.
Objectives: To investigate the possible association between corneal biomechanical parameters and morphological properties in a normal myopic population.Methods and patients: The study consisted of 480 normal myopic eyes (240 healthy volunteers), with ages ranging from 18 to 44 years (mean, std 23.84 ± 5.08 years), and mean spherical equivalent (MSE) ranging from -14.00 to -1.13 D (mean, std –5.68 ± 2.17 D). Corneal hysteresis (CH) and corneal resistance factor (CRF) were measured using the Ocular Response Analyser (ORA; Reichert Ophthalmic Instruments, Depew, New York, USA) in both eyes. Pentacam (Oculus GmbH, Wetzlar, Germany, and software version 1.17r27) were used to obtain corneal central elevation and corneal asphericity (Q value within 6 mm diameters) of both anterior and posterior surfaces, corneal central thickness (CCT), corneal volume (CV) and corneal spherical aberration values. Corneal volume was calculated within diameters from 1.0 to 6.0 mm with 0.5 mm steps centered on the apex to create the corneal-volume distribution. Pearson correlation coefficient was used for analysis on relationships between CH, CRF and Pentacam parameters (CV, elevation, Q-value, spherical aberrations, et al).Result: The values of CH and CRF presented normal distribution and the mean CH was (10.38 ± 1.36) mmHg, and the mean CRF was (10.70 ± 1.59) mmHg. There is a good correlation between CH, CRF and CCT (CH: r=0.54, P=0.000*, CRF: r=0.61, P=0.000*), and a stable correlation with each CV value (r≈0.5, P=0.000*) within central 6 mm diameter corneal region. On the other hand, CH and CRF were negatively correlated with anterior central elevation (CH: r=-0.136*, P=0.002*; CRF: r=-0.152*, P =0.001*), positively correlated with Q value of anterior surface (CH: r=0.136*, P=0.002; CRF: r=0.132*, P=0.003) and corneal spherical aberration (CH: r=0.184*, P=0.000*; CRF: r=0.191*, P =0.000*).Conclusions: There is a homogeneous relationship displayed between corneal biomechanical parameters (CH and CRF) and corneal morphological features. Our results suggest high biomechanical values might be related to central flattening and oblate corneal shape.
Subjects
Ophthalmology & OptometryAsk about this article
Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.
More from this journal
The newest in Journal of Clinical and Experimental Ophthalmology.
Short communication · 30 Aug 2026
Vision Science in the Modern Era: Integrating Neuroscience, Genetics, and Technology
Commentary · 3 Aug 2026
Emerging Frontiers in Experimental Ophthalmology: Innovations and Translational Challenges
Conferences in this field
Upcoming conferences in this field.
Follow
New articles from what you follow go into your email digest, once you agree to alert emails.
- JournalJournal of Clinical and Experimental Ophthalmology
- SubjectOphthalmology & Optometry
- BylineArticles whose byline prints “Lin Zhang”
- BylineArticles whose byline prints “Yan Wang”
- BylineArticles whose byline prints “Lili Xie”
- BylineArticles whose byline prints “Weili Geng”
- BylineArticles whose byline prints “Tong Zuo”
