The Hazardous Air Pollutants on Site (HAPSITE), a portable Gas Chromatograph-Mass Spectrometer (GCMS), has been used to detect, identify, and quantify Volatile Organic Compounds (VOCs) from environmental samples, providing on-site analysis to aid in operational risk management. HAPSITE is equipped with a hand-held sampling probe in which an air sample is delivered into a concentrator, and the VOCs collected in the concentrator are transferred, separated, and identified in the GC-MS. An upgraded version, HAPSITE ER, has recently been introduced with additional sampling capability for solid phase micro extraction and Thermal Desorption (TD). To our knowledge, however, no study has yet evaluated the performance of the thermal desorber accommodated in HAPSITE ER. In this study, therefore, we analyzed EPA Method TO-15 compounds with two different sampling methods (probe and thermal desorber for TD tubes) in a HAPSITE ER, and compared their results against each other. A major finding was that the peak intensities of the TO-15 compounds, particularly those with high Boiling Point (BP), were substantially higher in the results obtained with the thermal desorber than in those with the sampling probe. The lower peak intensities of the compounds observed in the probe analysis are likely due to the condensation of the VOCs in the probe (transfer) line that is 6 feet long and maintained at 40°C as they are delivered from the probe to the concentrator, whereas the thermal desorber is directly connected to the HAPSITE (no transfer line is used), thereby eliminating the condensation of VOCs. In conclusion, our study suggests that for the analysis of VOCs with high up to 220°C, the use of TD tubes followed by desorption in the thermal desorber offered by the newer version of HAPSITE is recommended.

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.

Try:

More from this journal

The newest in Journal of Analytical & Bioanalytical Techniques.

All its articles

Conferences in this field

Upcoming conferences in this field.

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalJournal of Analytical & Bioanalytical Techniques
  • SubjectBiochemistry & Biotechnology
  • SubjectChemistry
  • BylineArticles whose byline prints “Jae Kwak”
  • BylineArticles whose byline prints “Maomian Fan”
  • BylineArticles whose byline prints “Brain A Geier”
  • BylineArticles whose byline prints “Claude C Grigsby”
  • BylineArticles whose byline prints “Darrin K Ott”