Preliminary study in in vivo CT dosimetry using optically stimulated luminescence detector (OSLD)
Universitas Gadjah Mada, Indonesia
Abstract
Select any words in the title or abstract, or press this and choose a sentence.
Optically stimulated luminescence nanoDot dosimeters (OSLD) have been used to measure single point dose distribution along the superior-inferior position at the surface (skin) and in depths, analogous to organ depths. A customized acrylic phantom of 46 cm length, 20 cm diameter was fabricated, and used for the experimental measurements using CT (Computed Tomography) beam. The CT dose profile along the surface and the center of the phantom, as a function of scan field, were measured. The measured dose due to the primary radiation was quantified and compared with the theoretical dose profile using conventional 100 mm pencil CT ionization chamber. The OSLD responses at the center axial and the surface were found to be similar with the dose profile from CT ionization chamber. In this study, the dose at the center axial was found higher than the surface dose up to 28.13%. The dose at the diagonal center was found to be the highest compared to the surface dose up to 61.13%. The result of this work has revalidated the potential use of OSLD in CT Dosimetry as an alternative to TL-Dosimetry.
Subjects
Oncology & CancerAsk about this article
Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.
Follow
New articles from what you follow go into your email digest, once you agree to alert emails.
- JournalCancer Science & Therapy
- SubjectOncology & Cancer
- BylineArticles whose byline prints “Muhammad Yusuf”