Biomechanical Effect of an Interlaminar Device on Ranges of Motion,Intradiscal Pressure,and Centers of Rotation

Authors: Giancarlo Guizzardi, Sylvain Persohn, Sophie Campana, Caspar Aylott, Piero Petrini, Wafa Skalli

The Open Access Journal of Science and Technology

Volume 3 · Issue 5 · Pages 1–13

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

Introduction.The Intra SPINE is an e winter laminar device that has been proposed with the aim to decompress the spinal canal with out reducing the extension motion.The purpose of this study was therefore to evaluate the biomechanical behavior of L4-L5 spinal units implanted with this interlaminar device, in terms of ranges of motion, intradiscal pressure, and centers of rotation. Material and Methods. Six human lumbar spines were harvested within 10 days after death. A speciï¬Âc spine testing device was used to apply moments upto10Nminï¬Â‚exion-extension,lateral bending(left-rightï¬Â‚exion)and left-right axial rotation (torsion),with measurement of vertebral 3D motion and of intervertebrald is c pressure. Protocol was repeated for each specimen in 5 conï¬Âgurations: intact specimen; after L4-L5 bilateral medial hemifacetectomy and both yellow ligament resection; after implantation of the interlaminar device at the L4-L5 level; after removal of the L4-L5 supraspinous ligament, resection of the posterior third of the discand addition of an artiï¬Âcial ligament;after device and artiï¬Âcial ligament removal. Results.Theimplant reduced increases in segmental ï¬Â‚exion seen following injury particularly when applied with the artiï¬Âcial ligament. Intradiscal pressure reduced following application of the implant with out reducing extension range.A small posterior shift of the Mean Centers of Rotation(MCR) was noticed after instrumentation.To rsion and lateral bending range was unaï¬Â€ected by the interlaminar device. Conclusion.This biomechanical study yields a better understanding of this interlaminar implant eï¬Â€ect.A large clinical trial with follow-up would be required to evaluate and conï¬Ârm in vivo the observed invitro biomechanical behavior of the device

Subjects

Orthopaedics

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.

Try:

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalThe Open Access Journal of Science and Technology
  • SubjectOrthopaedics
  • BylineArticles whose byline prints “Giancarlo Guizzardi”
  • BylineArticles whose byline prints “Sylvain Persohn”
  • BylineArticles whose byline prints “Sophie Campana”
  • BylineArticles whose byline prints “Caspar Aylott”
  • BylineArticles whose byline prints “Piero Petrini”
  • BylineArticles whose byline prints “Wafa Skalli”