The results of a TLD Therapy Dosimetry Quality Assurance Program for Dong Nai General Hospital, Vietnam (Period 2013-2017)
Abstract
Select any words in the title or abstract, or press this and choose a sentence.
The objective of this study is to perform a thermo luminescent dosimetry (TLD) postal dose audit provided the International Atomic Energy Agency (IAEA) for a linear accelerator in Dong Nai General Hospital. The TLD were irradiated with an absorbed dose to water of 2 Gy for 6 MV and 15 MV photon beams generated by a linear accelerator (LINAC) (Siemens Medical Solutions, Concord, CA). The percentage deviation relative to IAEA measured dose was 0.4% and 3.4% for 6 MV and 16 MV, respectively. Agreement within ± 5% between the users stated dose and the IAEA measured dose is considered satisfactory. The results showed that all photon beams are within the IAEA’s acceptance limit of ± 5%. As a conclusion, the beam output of Siemens Primus is accuracy in delivery of absorbed dose within ± 5%.
Ask about this article
Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.
More from this journal
The newest in Nuclear Medicine & Radiation Therapy.
Brief report · 1 Dec 2025
AI: Precision Radionuclide Therapy Dosimetry Advancements
Brief report · 1 Dec 2025
Multi-Isotope Therapy: Powering Precision Nuclear Oncology
Follow
New articles from what you follow go into your email digest, once you agree to alert emails.
- JournalNuclear Medicine & Radiation Therapy
- SubjectRadiology & Imaging
- SubjectEnergy & Power
- BylineArticles whose byline prints “Duong Thanh Tai”
- BylineArticles whose byline prints “Truong Thiet Dung”