Article · 22 Sep 2026
Research articlePublished 3 May 2019
Simultaneous estimation and validation of Drotaverine Hydrochloride and Nimesulide in tablet dosage form using Reversed-Phase HPLC
Abstract
Select any words in the title or abstract, or press this and choose a sentence.
A simple, rapid, fast and precise reversed-phase high performance liquid chromatographic method has been developed and validated for the simultaneous estimation of drotaverine hydrochloride and nimesulide in tablet dosage form. Best separation was achieved using a Luna C18 (5μm×25cm×4.6mm i.d) column with the flow rate of 1.0 ml min-1. The analytes were monitored at 239 nm. The mobile phase consisted of acetonitrile - triethylamine – water (65:0.3:34.7 v/v/v), adjusted to pH 4.5 with 5% o-phosphoric acid. Under this conditions the retention time were of 4.14 and 5.87 minute respectively for drotaverine hydrochloride and nimesulide, indicating that the compounds were well separated. Total run was completed within eight minute. The method was validated for accuracy, precision, linearity, specificity and sensitivity as per ICH norms. The developed and validated method was successfully used for quantitative analysis of commercial formulation Nobel spas tablets. Validation study revealed that the method is specific, rapid, accurate, precise, reliable and reproducible.
Subjects
Pharmaceutical SciencesAsk about this article
Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.
More from this journal
The newest in Der Pharma Chemica.
Conferences in this field
Upcoming conferences in this field.
Follow
New articles from what you follow go into your email digest, once you agree to alert emails.
- JournalDer Pharma Chemica
- SubjectPharmaceutical Sciences
- BylineArticles whose byline prints “Raj K. Prasad”
- BylineArticles whose byline prints “Rajesh Sharma”