Assessment of PET scanner quantitative factors: Analysis and validation of GATE Monte Carlo simulation with experimental study

Authors: Mehrnoosh Karimipourfard, Sedighe Sina, M. Sadeghi, R. safari & M.S. Alavi

Imaging in Medicine

Volume 13 · Issue 9 · Pages 1–12

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

The quality control procedure would also be a requirement for any department to achieve a level of formal accreditation. The QC of the PET images was performed using NEMA and Jaszczak phantoms. This article describes the experimental and simulation quantitative protocols of acquiring PET imaging parameters to validate and analyze the result to use in complex geometry, patient dosimetry, and the optimal value of scanner performance in comparison studies. The QC factors such as signal-to-noise ratio, SNR, contrast to noise ratio, CNR, standardized uptake values, SUV, and spatial resolution were estimated from the hot and cold PET-CT images of the phantoms containing various activities. GATE Monte Carlo code was used for simulation of the phantoms. The images were reconstructed by STIR software, and the QC parameters of the simulation-based images were obtained. The maximum differences between SNR results of the Monte Carlo simulation, and the experimental results were found to be 2.29 and 2.93, and the maximum differences in CNR values were reported to be 2.79 and 2.2 for 20 and 40 MBq/ml activates respectively. The spatial resolution of the images was obtained by the size of spheres and rods of the Jaszczak phantom. According to the results 12.7 and 15.9 mm sphere diameters and 8 mm rod diameters were invisible in 20 and 40 MBq/ml cold scans. The simulation results show that the GATE Monte Carlo code is a beneficial tool for the simulation of PET images. These findings, while preliminary, suggest that analyzing the factors in patient dosimetry with a relation of the pixel intensity value of images with activity variation.

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Scholars Central does not store it.

Try:

More from this journal

The newest in Imaging in Medicine.

All its articles

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalImaging in Medicine
  • SubjectRadiology & Imaging
  • BylineArticles whose byline prints “Mehrnoosh Karimipourfard”
  • BylineArticles whose byline prints “Sedighe Sina”
  • BylineArticles whose byline prints “M. Sadeghi”
  • BylineArticles whose byline prints “R. safari & M.S. Alavi”